>
>
2026-09-04
Applicable to: 3300 XL series probes (8/11/14 mm) and 330180 series Proximitors used with 3500 vibration/displacement monitoring modules. The process involves five steps—ranging from simple to advanced—to quickly pinpoint faults: initial visual inspection → power-off electrical testing → powered voltage verification → professional calibration (using TK-3E) → 3500 system alarm verification.
Disconnect the probe from the extension cable and measure the resistance between the probe's BNC center pin and the shielded housing:
Fault diagnosis:
Measure between the probe's center pin and the metal housing/armored shield:
Insulation between the Proximitor's power input/signal output terminals and the housing must be ≥100 MΩ; low insulation indicates moisture ingress or internal circuit breakdown.
Disconnect the probe circuit and power up only the Proximitor. Measure the voltage between VT and COM; it should be stable between -18 and -24V DC. If there is no voltage, the voltage is too low, or the polarity is reversed, address the power supply issue first rather than assuming the sensor is damaged.
Disconnect the probe/extension cable. Use a metal object to short-circuit the BNC center pin to the outer shield housing on the Proximitor side:
Fault diagnosis: If the voltage falls outside this range, is absent, or tracks the supply voltage, the internal oscillation/demodulation circuit is damaged; replace the unit.
Align the probe with a clean carbon steel target surface and slowly advance it to the linear midpoint (standard zero-gap distance is approximately 1.27 mm / 50 mils):
Standard sensitivity for an 8mm probe is 7.87 V/mm (200 mV/mil); a voltage error of ≤ ±0.5% of full scale at each point is considered acceptable.
![]()
Fault Symptom | Likely Damaged Component
| Fault Symptom | Likely Damaged Component |
|---|---|
| Coil resistance infinite/0Ω | Internal probe wire break/short circuit |
| Insulation resistance extremely low | Probe/cable moisture ingress or damaged insulation |
| Output ≠ -0.6 to -0.8 V after shorting BNC | Preamplifier damaged |
| Gap voltage shows constant value instead of smooth change | Cable open/short circuit |
| TK-3E linearity/sensitivity significantly out of tolerance | Probe aging or proximitor drift |
| 3500 channel shows continuous "Probe Fault" (red light) | Loop open/short circuit; locate fault by measuring resistance in sections |
⚠️ Key Precautions
Contact Us at Any Time